Title of article :
Preliminary study to optimize the irradiation condition for future application in small animal CT
Author/Authors :
Bisogni، نويسنده , , M.G. and Delogu، نويسنده , , P. and Carpentieri، نويسنده , , C. and Fantacci، نويسنده , , M.E. and Quattrocchi، نويسنده , , M. and Rosso، نويسنده , , V. and Stefanini، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
4
From page :
142
To page :
145
Abstract :
In digital imaging the ability to detect low contrast details is an aim that must be pursued. In order to achieve this result we have developed an acquisition system based on silicon pixel detector. The detector is a matrix of 256×256 square pixels of 55 μm size. It is bump-bonded to a readout chip developed by the Medipix2 Collaboration [X. Llopart, et al., IEEE Trans. Nucl. Sci, NS-49 (5) (2002) 2279]. To see the low contrast details with this system, it is necessary to reduce inside the image the fluctuations of pixel-to-pixel counts, [1] down to statistical fluctuations. This has been achieved equalizing the phantom image with a high statistics flat field acquired in the same irradiation condition. We have studied the dependence of the image uniformity for different flat field and we have investigated the dependence of the image quality on the thickness of the phantom and on the energy window selected inside the incident spectrum.
Keywords :
image quality , Solid-state detector , digital imaging
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2006
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2200956
Link To Document :
بازگشت