Title of article :
A large tracking detector for low-energy ions
Author/Authors :
Bougamont، نويسنده , , E. and Bourgeois، نويسنده , , P. and Drouart، نويسنده , , A. and Mazur، نويسنده , , C. and Nalpas، نويسنده , , L. and Pollacco، نويسنده , , E.C and Riallot، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
3
From page :
129
To page :
131
Abstract :
We present an innovative detector with a large working area (100×400 mm2) for ion tracking. This device is based on the detection of secondary electron emitted by a thin foil placed at 45° to the direction of the beam. Secondary electrons are accelerated and focused by an electric and a magnetic field toward a low-pressure gaseous chamber. Time and position (X and Y) are measured. The typical resolutions (FWHM) are 300 ps and 1.5 mm, respectively.
Keywords :
Secondary electrons , Low pressure , Gaseous detector , Radioactive beam , Spatial resolution
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2004
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2201068
Link To Document :
بازگشت