Title of article :
Experience with a factory-calibrated HPGe detector
Author/Authors :
Bossus، نويسنده , , D.A.W. and Swagten، نويسنده , , J.J.J.M. and Kleinjans، نويسنده , , P.A.M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
650
To page :
654
Abstract :
For k0-based analysis, an HPGe detector has to be used. This detector has to be absolutely calibrated in a reference position and with a defined geometry so that, using SOLCOI/KAYZERO software package, for example, efficiencies of other positions and sample geometries can be calculated. This reference calibration is a time-consuming procedure during which the detector is not available for analyses. Therefore, DSM Resolve decided to purchase a “factory-calibrated” detector. Efficiency calibrations were ordered for a point-source geometry at a coincidence-free distance from the detector and for two additional distances closer to the detector. After delivery, the factory calibration was checked at DSM Resolve using a limited set of PTB-calibrated reference sources. At the end, we decided nevertheless to perform a standard and full calibration of the detector, because it turned out that the factory-calibrated detector was not accurate enough to be used for quantitative analyses.
Keywords :
k0-NAA , Ge detector , Full-energy peak efficiency , KAYZERO , SOLCOI , LabSOCS
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2006