Title of article :
Energy scales in X-ray microcalorimeters
Author/Authors :
Tillotson، نويسنده , , W.A and Boyce، نويسنده , , K.R. and Brown، نويسنده , , G.V. and Cottam، نويسنده , , J and Figueroa-Feliciano، نويسنده , , E and Kelley، نويسنده , , R.L and Porter، نويسنده , , F.S and Stahle، نويسنده , , C.K، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
4
From page :
595
To page :
598
Abstract :
Microcalorimeter pulse shape characteristics, such as pulse height, decay constant, and rise time, are dependent on the detector temperature and bias as well as incident photon energy. We examine the nature of the temperature dependency by illuminating the Astro-E2 X-ray Spectrometer microcalorimeter array with X-rays generated by electron impact on a selection of foil targets. The resulting pulses are collected over a range of detector temperatures. We observe and model these pulses by fitting the data with non-linear pulse models. Our aim is to determine a robust method for correcting the energy scale obtained in ground calibration for slight differences in the operating conditions while in orbit.
Keywords :
Thermal detectors , Microcalorimeter array
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2004
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2202077
Link To Document :
بازگشت