Title of article :
A low-angle X-ray scatter breast-imaging system based on a novel multi-element Si-pad array operated in spectroscopic mode
Author/Authors :
Darambara، نويسنده , , D.G and Sellin، نويسنده , , P.J and Maehlum، نويسنده , , G، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
A novel multi-element Si-pad array operated in spectroscopic mode was designed in collaboration with Sintef to measure low-angle X-ray scattered photons in the energy range of 12–45 keV with an FWHM energy resolution of about 4 keV. This detector is used as the ultra-sensitive diffraction detector in a dual-sensor X-ray breast-imaging scanning system. Each detector is wire-bonded to a high-density low-noise XA1.3 readout chip designed by IDE AS. A new DAQ system was also developed to be used in conjunction with the detector and the readout electronics. Results regarding the characterization of this new Si-pad detector and its performance to measure scattering signatures of several materials of interest are presented.
Keywords :
x-ray imaging , X-ray detectors , Si-pad detectors , Readout electronics , X-ray diffraction , Low-angle X-ray scattering
Journal title :
Nuclear Instruments and Methods in Physics Research Section A