Title of article
Industrial X-ray imaging based on scintillators and CMOS APS array: direct X-ray irradiation effects
Author/Authors
Kim، نويسنده , , Kwang Hyun and Jeon، نويسنده , , Sung Chae and Kim، نويسنده , , Young Soo and Cho، نويسنده , , Gyuseong، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
4
From page
454
To page
457
Abstract
To see the effects of the direct X-ray in a Lanex screen-coupled CMOS APS imager, we measured modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). These measurements were performed under the condition of non-destructive test (NDT). By increasing the cumulative exposure on the imager, the MTF was degraded, and also leading to the DQE degradation. Each parameter changed by the exposure is described in detail.
Keywords
CMOS APS , DQE , MTF , NDT , Direct X-ray
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2005
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2203030
Link To Document