• Title of article

    Response function calculation of an underwater gamma ray NaI(Tl) spectrometer

  • Author/Authors

    Vlachos، نويسنده , , D.S. and Tsabaris، نويسنده , , C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    7
  • From page
    414
  • To page
    420
  • Abstract
    Autonomous measuring systems for radioactivity in the water environment require caution concerning the specifications of power consumption, stability, communication equipment, tolerance and efficiency. In addition, the calculation of the response function of the whole system is essential for suppressing the background of the measurement and for estimating the concentration of the involved radionuclides, especially given the greater probability of primary gamma photons undergoing multiple scattering events before they interact with the sensing device. In this work, a method is presented that can be used to calculate the response function of a NaI(Tl)-based spectrometer when it is used in the marine environment. The method takes into account both the scattering of photons in the water (analytical calculations) and the detection mechanism of the sensor (Monte-Carlo simulation). In order to validate the method, the calculated response function has been used in a real measuring system in order to estimate the concentration of 40K and thus the salinity of the water.
  • Keywords
    NaI(Tl) detector , response function , Marine radioactivity
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2005
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2203187