Title of article
CYBPET: a cylindrical PET system for breast imaging
Author/Authors
Karimian، نويسنده , , A. and Thompson، نويسنده , , C.J. and Sarkar، نويسنده , , S. and Raisali، نويسنده , , G. and Pani، نويسنده , , R. and Davilu، نويسنده , , H. and Sardari، نويسنده , , D.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
9
From page
427
To page
435
Abstract
We propose a Cylindrical Breast PET (CYBPET) system for breast imaging with patients in the prone position. An individual pendulous breast is covered by thin plastic to provide reduced pressure fixation and surrounded by the crystals inside the CYBPET ring. Each breast is imaged separately. The rest of the body is shielded properly to minimize the contribution of scattered photons from the other breast and the rest of the body. To compare the CYBPET with whole-body PET (WB-PET) the simulations of CYBPET and a WB-PET (GE-Advance) for a 10 mm tumor inside the breast with a lesion to background (breast) activity concentration of 6 to 1 were made. The noise effective count rate (NECR) of CYBPET is about twice that of WB-PET at activity concentrations less than 3.1 μCi/cc. The spatial resolution of CYBPET is better by 25% than the WB-PET.
Keywords
Positron emission mammography , Nuclear medicine instrumentation , breast imaging , breast cancer , PEM , Positron emission tomography
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2005
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2203643
Link To Document