Title of article
Apparatus for X-ray diffraction microscopy and tomography of cryo specimens
Author/Authors
Beetz، نويسنده , , T. and Howells، نويسنده , , M.R. and Jacobsen، نويسنده , , C. and Kao، نويسنده , , C.-C. and Kirz، نويسنده , , J. and Lima، نويسنده , , E. and Mentes، نويسنده , , T.O. and Miao، نويسنده , , H. and Sanchez-Hanke، نويسنده , , C. and Sayre، نويسنده , , D. and Shapiro، نويسنده , , D.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
10
From page
459
To page
468
Abstract
An apparatus for diffraction microscopy of biological and materials science specimens is described. In this system, a coherent soft X-ray beam is selected with a pinhole, and the illuminated specimen is followed by an adjustable beamstop and CCD camera to record diffraction data from non-crystalline specimens. In addition, a Fresnel zone plate can be inserted to allow for direct imaging. The system makes use of a cryogenic specimen holder with cryotransfer capabilities to allow frozen hydrated specimens to be loaded. The specimen can be tilted over a range of ± 80 ∘ for three-dimensional imaging; this is done by computer-controlled motors, enabling automated alignment of the specimen through a tilt series. The system is now in use for experiments in soft X-ray diffraction microscopy.
Keywords
X-ray microscopy , x-ray imaging , X-ray tomography , Diffraction microscopy , lensless imaging