Title of article :
Electrical characteristics of hybrid detector based Gd2O2S:Tb-Selenium for digital radiation imaging
Author/Authors :
Kang، نويسنده , , Sang-Sik and Park، نويسنده , , Jikoon and Choi، نويسنده , , Jang-Yong and Cha، نويسنده , , Byung-Yul and Cho، نويسنده , , Sung-Ho and Nam، نويسنده , , Sang-Hee، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Fine Gd2O2S:Tb powders were synthesized by using a solution-combustion method for a high-resolution digital X-ray imaging detector. The PL spectrum showed that the phosphor was fully crystallized and that the Tb3+ ions substituted well for the Gd3+ sites. To investigate the X-ray response of the phosphor, a uniform Gd2O2S:Tb film was grown using a screen-printing method. The X-ray sensitivities of the 100 μm-Gd2O2S:Tb/30 μm -Se and 200 μm -Se detector were 470 and 420 pC/cm2/mR, respectively, at an electric field of 10 V/μm. The results of the study suggest that the hybrid detector has a significant potential in the application of digital radiography and fluoroscopy systems.
Keywords :
Digital X-ray detector , Gd2O2S:Tb , X-ray sensitivity , Amorphous selenium
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A