Title of article
Development of scintillator for a high-frame-rate neutron radiography
Author/Authors
Matsubayashi، نويسنده , , Masahito and Katagiri، نويسنده , , Masaki، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
5
From page
389
To page
393
Abstract
The properties of Ni doped ZnS(Ag) scintillator for a high-frame-rate neutron radiography and also for a high-counting-rate neutron scintillation detector were examined and confirmed to be promising. Although the deterioration of the emission spectrum and in the light transmission property was observed, a slow component in the scintillation decay was well suppressed. The decrease in the thermal neutron detection efficiency such as small percent due to the deterioration of optical property was recoverable with the replacement of neutron converter such as 6LiF to 10B2O3.
Keywords
ZnS(Ag):Ni scintillator , Neutron radiography , Scintillation decay , scintillation detector
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2004
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2203741
Link To Document