• Title of article

    Ion beam induced charge imaging of epitaxial GaN detectors

  • Author/Authors

    Sellin، نويسنده , , P.J. and Hoxley، نويسنده , , D. and Lohstroh، نويسنده , , A. and Simon، نويسنده , , A. and Cunningham، نويسنده , , W. and Rahman، نويسنده , , M. and Vaitkus، نويسنده , , J. and Gaubas، نويسنده , , E.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    5
  • From page
    82
  • To page
    86
  • Abstract
    We report the use of ion beam induced charge imaging to characterise the charge signal uniformity of epitaxial gallium nitride radiation detectors. The detectors were fabricated from 2 μm thick semi-insulating gallium nitride, grown by MOCVD on a sapphire substrate. A carrier concentration of 1.4×1015 cm−3 was measured using capacitance–voltage measurements. Ion beam induced charge imaging was carried out with a 2 MeV alpha particle beam focussed to a 3 μm diameter and raster scanned across the device. The resulting ion beam images show excellent charge signal uniformity in this material with no evidence of material defects or polycrystalline structure on the micrometer length scale. No evidence of charge signal trapping was observed in these devices.
  • Keywords
    Gallium nitride , radiation detector , Ion beam induced charge imaging
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2004
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number