• Title of article

    3D position sensitivity of a highly segmented Ge detector

  • Author/Authors

    Milechina، نويسنده , , L. and Cederwall، نويسنده , , B.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    14
  • From page
    278
  • To page
    291
  • Abstract
    The 3D position resolution of a planar HPGe detector with a 25-fold segmented cathode contact has been studied. The measurements were performed with a collimated 137Cs γ -ray pencil beam and a coincidence scattering technique that selects single interaction events at well-defined irradiation points. The γ rays scattered at 90 ∘ , depositing 374 keV energy in the detector, were registered and the pulse shapes from the irradiated segment and three neighboring segments were analyzed. The resulting average 3D position sensitivity for single γ -ray interactions is estimated at 0.4 mm
  • Keywords
    Position sensitivity , Pulse shape analysis , Segmented planar germanium detectors