Title of article :
Towards a continuous crystal APD-based PET detector design
Author/Authors :
Bruyndonckx، نويسنده , , Peter and Lemaître، نويسنده , , Cedric and Schaart، نويسنده , , Dennis and Maas، نويسنده , , Marnix and van der Laan، نويسنده , , D.J. (jan) and Krieguer، نويسنده , , Magalie and Devroede، نويسنده , , Olivier and Tavernier، نويسنده , , Stefaan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Monolithic scintillator blocks provide an efficient way to significantly increase the sensitivity of high-resolution positron emission tomography (PET) systems. Using statistical or neural network-based positioning schemes, the measured intrinsic detector resolution for perpendicular incident photons is 1.7 mm full-width at half-maximum (FWHM) for 10 mm thick LSO blocks and 2.0 mm FWHM for 20 mm thick LSO blocks. Because the positioning algorithms determine the incidence position as opposed to the combination of interaction position and interaction depth, a very good parallax correction can be achieved. Energy and detector time resolution were 11.5% FWHM and 1.6 ns FWHM, respectively.
Keywords :
Positron emission tomography , APD , Scnitillator , neural network , Depth of interaction , Statistical positioning algorithm