• Title of article

    Towards a continuous crystal APD-based PET detector design

  • Author/Authors

    Bruyndonckx، نويسنده , , Peter and Lemaître، نويسنده , , Cedric and Schaart، نويسنده , , Dennis and Maas، نويسنده , , Marnix and van der Laan، نويسنده , , D.J. (jan) and Krieguer، نويسنده , , Magalie and Devroede، نويسنده , , Olivier and Tavernier، نويسنده , , Stefaan، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    182
  • To page
    186
  • Abstract
    Monolithic scintillator blocks provide an efficient way to significantly increase the sensitivity of high-resolution positron emission tomography (PET) systems. Using statistical or neural network-based positioning schemes, the measured intrinsic detector resolution for perpendicular incident photons is 1.7 mm full-width at half-maximum (FWHM) for 10 mm thick LSO blocks and 2.0 mm FWHM for 20 mm thick LSO blocks. Because the positioning algorithms determine the incidence position as opposed to the combination of interaction position and interaction depth, a very good parallax correction can be achieved. Energy and detector time resolution were 11.5% FWHM and 1.6 ns FWHM, respectively.
  • Keywords
    Positron emission tomography , APD , Scnitillator , neural network , Depth of interaction , Statistical positioning algorithm
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number