Title of article :
Contrast noise behaviour of a rotating slat collimated gamma camera
Author/Authors :
Van Holen، نويسنده , , R. E. Vandenberghe، نويسنده , , S. and Staelens، نويسنده , , S. and D’Asseler، نويسنده , , Y. and Lemahieu، نويسنده , , I.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
274
To page :
277
Abstract :
Compared to classical parallel hole (PH) collimated gamma cameras, the geometry of a rotating slat (RS) collimated detector implies a better sensitivity versus resolution trade-off. Nevertheless, the expected improvement in image quality from this device is degraded due to the assumptions made during the image reconstruction of planar images. This study shows that an accurate model of the imaging process for iterative reconstruction is necessary in order to improve planar image quality. To this end, a Monte Carlo based system matrix was derived from a simple simulation by exploiting the rotational and translational symmetries present in this acquisition set-up. It was found that this system matrix improves the contrast and noise characteristics of reconstructed images compared to a reconstruction with no modelling. As a result, the RS device is able to yield improved quality for planar images compared to a PH camera.
Keywords :
Monte carlo based reconstruction , Rotating slat collimator
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2204726
Link To Document :
بازگشت