Title of article
Characteristics of optical imaging capillary plate gas detector with fiber optic plate
Author/Authors
Tokanai، نويسنده , , Fuyuki and Sakurai، نويسنده , , Hirohisa and Gunji، نويسنده , , Shuichi and Kaneko، نويسنده , , Masahiro and Endo، نويسنده , , Tetsuro and Fujita، نويسنده , , Yoshio and Atsumi، نويسنده , , Takuji and Okada، نويسنده , , Teruyuki، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
5
From page
289
To page
293
Abstract
A capillary plate (CP) gas detector with fiber optic plate (FOP) has been successfully operated as a gas scintillation proportional counter. A thin film of indium tinoxide (ITO) was used for the transparent electrode of the CP. Scintillation light emitted during the development of electron avalanches can be observed through the FOP using a photomultiplier tube (PMT) and an optical imaging system, which consists of a CP gas detector and a cooled CCD camera coupled to lens optics. Using this optical imaging system, clearer images of X-rays have been obtained. Successful operation with the CP gas detector and the FOP has allowed us to realize a novel imaging device for medical imaging, cellular function analysis, and X-ray imaging.
Keywords
CCD , microstructure , Scintillation , gas mixtures , Capillary plate
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2007
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2204733