Title of article :
Analysis of the expected AGIPD detector performance parameters for the European X-ray free electron laser
Author/Authors :
Potdevin، نويسنده , , Guillaume and Graafsma، نويسنده , , Heinz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
8
From page :
229
To page :
236
Abstract :
A review of the parameters of the AGIPD detector planned to be used on the SASE 1 and SASE 2 beamlines of the European XFEL is presented. For both Coherent X-ray Diffraction Imaging (CXDI) and X-ray Photon-Correlation Spectroscopy (XPCS), the required detector noise level and dynamic range are evaluated from signal and background simulations on a per instrument basis. A worst case scenario of accumulated dose is computed in order to evaluate the required detector radiation hardness. Finally, acknowledging a mismatch between pixel size requirement and technological feasibility in the case of XPCS, the possibility to aperture the pixel in order to match the size requirement is presented and discussed.
Keywords :
Detector modelling and simulations , Instrumentation for FEL
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2011
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2205265