• Title of article

    Noise and trigger efficiency characterization of cooled silicon pad detectors

  • Author/Authors

    Marques، نويسنده , , A.C. and Wahl، نويسنده , , U. and Correia، نويسنده , , J.G. and Silva، نويسنده , , M.R. and Rudge، نويسنده , , A. and Weilhammer، نويسنده , , P. and Soares، نويسنده , , J.C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    9
  • From page
    1056
  • To page
    1064
  • Abstract
    Technical progress on silicon pad electron detectors, currently used in emission channelling experiments to study lattice location of radioactive dopants and impurities in single crystals, is reported. Noise and trigger efficiency improvements are achieved by using 500 μm and 1 mm thick detectors coupled to a cooled readout system. The static properties, noise, gamma ray and electron trigger efficiency and energy resolution for different temperatures under air and vacuum were measured. The advantages of the future implementation of 1 mm silicon pad detectors with cooled self-triggering readout chips are discussed.
  • Keywords
    Position-sensitive detectors , Cooling silicon pad detectors , Electron detection , Noise and trigger efficiency characterization
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2205424