Title of article :
Multiple occupancy considerations for the SmartPET imaging system
Author/Authors :
Beveridge، نويسنده , , Toby E. and Gillam، نويسنده , , John E. and Boston، نويسنده , , Andrew J. and Boston، نويسنده , , Helen C. and Cooper، نويسنده , , Reynold J. and Mather، نويسنده , , Andrew R. and Hall، نويسنده , , Chris J. and Nolan، نويسنده , , Paul J. and Lewis، نويسنده , , Robert A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
68
To page :
71
Abstract :
The SmartPET collaboration is investigating the use of two planar high-purity germanium double-sided strip detectors as a Compton imaging positron emission tomography system. Monte Carlo simulations suggest that a large proportion of interactions within the detectors will occur within a small spatial volume, introducing significant ambiguities within the position and energy measurements made by the detectors. Under certain circumstances, the ambiguities will result in multiple interactions being detected as a single interaction. This investigation studied the effect of such multiple occupancies and found that approximately 45% of 511 keV events include multiple interactions within a 5×5×20 mm3 volume. Nevertheless, the effect on the quality of the final data remains quite acceptable.
Keywords :
Compton imaging , Orthogonal strip detector , Positron emission tomography
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2205479
Link To Document :
بازگشت