Title of article :
Short induction gap gas electron multiplier (GEM) for X-ray spectroscopy
Author/Authors :
Mir، نويسنده , , J.A. and Stephenson، نويسنده , , R. and Rhodes، نويسنده , , N.J and Schooneveld، نويسنده , , E.M. and Veloso، نويسنده , , J.F.C.A. and dos Santos، نويسنده , , J.M.F.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
179
To page :
182
Abstract :
Experimental work was carried out to evaluate the performance of a Gas Electron Multiplier (GEM) operated with a micromesh readout plane that enabled the induction gap to be set at 50 μm. We measured the essential operational parameters of this system using Ar(75%)-isobutane (25%) as the counter gas mixture. The measurements included the effective gain, effective gain stability, and the X-ray energy resolution using a 5.89 keV X-ray source. These studies demonstrated several advantages of the current system when compared with the standard operation, such as lower operational voltages, higher effective gains and improved effective gain stability.
Keywords :
Gas electron multiplier , Effective gain , GEM , X-ray energy resolution