• Title of article

    Development of a soft X-ray pulse height analyzer on the J-TEXT tokamak

  • Author/Authors

    Jin، نويسنده , , W. and Zhuang، نويسنده , , G. and Ding، نويسنده , , Y.H. and Zhang، نويسنده , , Y. and Zhang، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    5
  • From page
    15
  • To page
    19
  • Abstract
    A soft X-ray pulse height analyzer has been designed and installed on the J-TEXT tokamak for measuring the electron temperature and possibly for monitoring the impurity content and concentration. A multi-channel analyzer, a set of amplifiers, and three silicon drift detectors constitute the detection system. An off-line analysis code was developed to analyze the raw data recorded by the system. The system can measure a minimum electron temperature of 400 eV, and its routine operation provides electron temperature measurements at three different positions with a temporal resolution of 60 ms and channel separation of 18 cm. Line radiation from some moderate and high Z impurities are observable by the system. The results show that the intensity of line radiation from heavy impurity ions is inversely proportional to the plasma density under similar discharge conditions.
  • Keywords
    pulse height analysis , Line radiation , Soft X-ray , tokamak
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2012
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number