Title of article :
Characterization of the composite bolometer with a high-Tc superconductor thermometer for an absolute radiometer of synchrotron radiation
Author/Authors :
Ivanov، نويسنده , , K.V. and Khokhlov، نويسنده , , D.A. and Khrebtov، نويسنده , , I.A. and Kulikov، نويسنده , , Yu.V. and Malyarov، نويسنده , , V.G. and Nikolenko، نويسنده , , A.D and Pindyurin، نويسنده , , V.F. and Zerov، نويسنده , , V.Yu.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
4
From page :
272
To page :
275
Abstract :
The article describes the structure and calculation modeling of characteristics of the detector for an absolute synchrotron radiation radiometer in the soft X-ray range of 8–0.4 nm. The detector is a bolometer working on the principle of substitution of electrical power for radiation power, with a high-temperature superconductor film as an element, sensitive to heat. The noise-equivalent power is predicted to be down to 4×10−11 W/Hz1/2 at 10 Hz.
Keywords :
Synchrotron radiation , Bolometer , high-temperature superconductor , MODELING
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2007
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2205895
Link To Document :
بازگشت