• Title of article

    Measurement of plasma parameters of an ion source for EAST-NBI using water-cooled Langmuir probe

  • Author/Authors

    Xie، نويسنده , , Yahong and Hu، نويسنده , , Chundong and Liu، نويسنده , , Sheng and Li، نويسنده , , Jun and Jiang، نويسنده , , Caichao، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    3
  • From page
    18
  • To page
    20
  • Abstract
    As the langmuir probe is a famous diagnosis tool for plasmas, the langmuir probe with cooling water is developed for an EAST-NBI high current ion source. Using the sawtooth wave sweeping power, the electron temperature and density are measured for the first time, and the plasma uniformity is analyzed with three langmuir probes. They are very useful to study the characteristics of high current ion sources.
  • Keywords
    Langmuir Probe , arc discharge , electron density , High current ion source
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2012
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number