Title of article
The effect of protons on the performance of second generation Swept Charge Devices
Author/Authors
Gow، نويسنده , , Jason P.D. and Holland، نويسنده , , Andrew D. and Pool، نويسنده , , Peter J. and Smith، نويسنده , , David R.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
4
From page
86
To page
89
Abstract
The e2v technologies Swept Charge Device (SCD) was developed as a large area detector for X-ray Fluorescence (XRF) analysis, achieving near Fano-limited spectroscopy at −15 °C. The SCD was flown in the XRF instruments onboard the European Space Agencyʹs SMART-1 and the Indian Space Research Organisationʹs Chandrayaan-1 lunar missions. The second generation SCD, proposed for use in the soft X-ray Spectrometer on the Chandrayaan-2 lunar orbiter and the soft X-ray imager on Chinaʹs HXMT mission, was developed, in part, using the findings of the radiation damage studies performed for the Chandrayaan-1 X-ray Spectrometer. This paper discusses the factor of two improvements in radiation tolerance achieved in the second generation SCD, the different SCD sizes produced and their advantages for future XRF instruments, for example through reduced shielding mass or higher operating temperatures.
Keywords
Proton radiation effects , X-ray fluorescence , CCD , Swept charge device
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2012
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2206035
Link To Document