• Title of article

    Estimation of dosimetry parameters for an EB accelerator using graphite calorimeters of different thickness

  • Author/Authors

    Benny، نويسنده , , P.G. and Khader، نويسنده , , S.A. and Sarma، نويسنده , , K.S.S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    4
  • From page
    108
  • To page
    111
  • Abstract
    Graphite calorimeters of different thickness in the range of 0.6 mm to 10 mm have been designed and fabricated for the dosimetry of 2 MeV electron beam accelerator. Average absorbed dose in each of the calorimeters of different thickness has been determined. The paper reports a method for selecting calorimeters with suitable thickness for its application as absorbed dose calorimeters and as total energy absorption calorimeters for an electron beam of particular energy. Also it reports, using calorimeters of different thickness, it is possible to estimate various parameters such as energy fluence, average absorbed dose, absorbed dose at any depth in the medium and practical range.
  • Keywords
    Electron beam accelerator , Dosimetry parameters , Graphite calorimeters
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2012
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number