Title of article :
High count rate spectroscopy with LaBr3:Ce scintillation detectors
Author/Authors :
Lِher، نويسنده , , B. and Savran، نويسنده , , Grasiela D. and Fiori، نويسنده , , E. and Miklavec، نويسنده , , M. and Pietralla، نويسنده , , N. and Vencelj، نويسنده , , M.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2012
Pages :
6
From page :
1
To page :
6
Abstract :
The applicability of LaBr3:Ce detectors for high count rate γ - ray spectroscopy is investigated. A 3 in.×3 in. LaBr3:Ce detector is used in a test setup with radioactive sources to study the dependence of energy resolution and photo peak efficiency on the overall count rate in the detector. Digitized traces were recorded using a 500 MHz FADC and analysed with digital signal processing methods. Good performance is obtained using standard techniques up to about 500 kHz counting rate. A pile-up correction method is applied to the data in order to further improve the capabilities at even higher rates with a focus on recovering the losses in efficiency due to signal pile-up. It is shown that γ - ray spectroscopy can be performed with only moderate lossen in efficiency and high resolution at count rates even above 1 MHz and that the performance can be enhanced in the region between 500 kHz and 10 MHz by using the applied pile-up correction techniques.
Keywords :
Digital pulse shape analysis , ? - ray spectroscopy , Lanthanum bromide , scintillator
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2012
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2206194
Link To Document :
بازگشت