    AX-PET: A novel PET concept with G-APD readout

  • Author/Authors

    Bolle، نويسنده , , E. and Casella، نويسنده , , C. and Chesi، نويسنده , , Keili E. and De Leo، نويسنده , , R. and Dissertori، نويسنده , , G. and Fanti، نويسنده , , V. and Gillam، نويسنده , , J.E. and Heller، نويسنده , , M. and Joram، نويسنده , , C. and Lustermann، نويسنده , , W. and Nappi، نويسنده , , E. and Oliver، نويسنده , , J.F. and Pauss، نويسنده , , F. and Rafecas، نويسنده , , M. and Rudge، نويسنده , , A. and Ruotsalainen، نويسنده , , U. and Sc، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    6
  • From page
    129
  • To page
    134
  • Abstract
    The AX-PET collaboration has developed a novel concept for high resolution PET imaging to overcome some of the performance limitations of classical PET cameras, in particular the compromise between spatial resolution and sensitivity introduced by the parallax error. tector consists of an arrangement of long LYSO scintillating crystals axially oriented around the field of view together with arrays of wave length shifter strips orthogonal to the crystals. This matrix allows a precise 3D measurement of the photon interaction point. This is valid both for photoelectric absorption at 511 keV and for Compton scattering down to deposited energies of about 100 keV. Crystals and WLS strips are individually read out using Geiger-mode Avalanche Photo Diodes (G-APDs). The sensitivity of such a detector can be adjusted by changing the number of layers and the resolution is defined by the crystal and strip dimensions. -PET modules were built and fully characterized in dedicated test set-ups at CERN, with point-like 22Na sources. Their performance in terms of energy ( R energy ≈ 11.8 % (FWMH) at 511 keV) and spatial resolution was assessed ( σ axial ≈ 0.65 mm ), both individually and for the two modules in coincidence. ampaigns at ETH Zurich and at the company AAA allowed the tomographic reconstructions of more complex phantoms validating the 3D reconstruction algorithms. ncept of the AX-PET modules will be presented together with some characterization results. We describe a count rate model which allows to optimize the planing of the tomographic scans.
  • Keywords
    Geiger-mode Avalanche Photo Diodes (G-APD) , SiPM , PET , Axial geometry
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2012
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2206576