Title of article :
Validation of k0 standardization method in neutron activation analysis—The use of Kayzero for Windows programme at the nuclear physics institute, Řež
Author/Authors :
Kube?ov?، نويسنده , , Marie and Ku?era، نويسنده , , Jan، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
4
From page :
403
To page :
406
Abstract :
The goal of this work was to validate the method of the k0 standardization in neutron activation analysis at the laboratory of neutron activation analysis in Řež using the Kayzero for Windows software package. For verification of the precision and accuracy of the method two types of reference materials were used. One was a set of synthetic materials aimed at checking the implementation of k0 standardization. The others were matrix standard reference materials of the US National Institute of Standards and Technology comprising several types of matrices. In general, good agreement was obtained between results of this work and values in certificates of the individual reference materials, thus proving the accuracy of our results and successful implementation of the method for analysis of real samples. In addition, detection limits for more than 50 elements were evaluated.
Keywords :
Validation , k0 standardization , NEUTRON ACTIVATION ANALYSIS
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2010
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2206782
Link To Document :
بازگشت