• Title of article

    Monte Carlo and least-squares methods applied in unfolding of X-ray spectra measured with cadmium telluride detectors

  • Author/Authors

    Moralles، نويسنده , , M. and Bonifلcio، نويسنده , , D.A.B. and Bottaro، نويسنده , , M. and Pereira، نويسنده , , M.A.G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2007
  • Pages
    4
  • From page
    270
  • To page
    273
  • Abstract
    Spectra of calibration sources and X-ray beams were measured with a cadmium telluride (CdTe) detector. The response function of the detector was simulated using the GEANT4 Monte Carlo toolkit. Trapping of charge carriers were taken into account using the Hecht equation in the active zone of the CdTe crystal associated with a continuous function to produce drop of charge collection efficiency near the metallic contacts and borders. The rise time discrimination is approximated by a cut in the depth of the interaction relative to cathode and corrections that depend on the pulse amplitude. The least-squares method with truncation was employed to unfold X-ray spectra typically used in medical diagnostics and the results were compared with reference data.
  • Keywords
    CdTe detector , diagnostic x-ray , Monte Carlo , GEANT4
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2007