Title of article
High-rate operation of the Micro-Hole and Strip Plate gas detector
Author/Authors
Veloso، نويسنده , , J.F.C.A. and Caldas، نويسنده , , Paula C.C. and Oliveira، نويسنده , , C.A.B. and Azevedo، نويسنده , , C.D.R. and dos Santos، نويسنده , , J.M.F. and Breskin، نويسنده , , A. and Chechik، نويسنده , , R.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
4
From page
362
To page
365
Abstract
We report the performance of the Micro-Hole-and-Strip-Plate (MHSP) operated in xenon, under high-flux irradiation of a 30-kV X-ray Cu-tube. Stable operation in pulse-counting mode for gas gains above 104 was measured up to counting rates of 5×105 Hz/mm2. A gain variation of less than 5% was observed for gains of 104, while the energy resolution for 8-keV X-rays degrades from ∼16% to ∼18%. No visible degradation was observed on the MHSP irradiated area.
Keywords
MHSP , Count rate , Xenon