Title of article :
Low-dose multiple-information retrieval algorithm for X-ray grating-based imaging
Author/Authors :
Wang، نويسنده , , Zhentian and Huang، نويسنده , , Zhifeng and Chen، نويسنده , , Zhiqiang and Zhang، نويسنده , , Li and Jiang، نويسنده , , Xiaolei and Kang، نويسنده , , Kejun and Yin، نويسنده , , Hongxia and Wang، نويسنده , , Zhenchang and Stampanoni، نويسنده , , Marco، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
5
From page :
103
To page :
107
Abstract :
The present work proposes a low dose information retrieval algorithm for X-ray grating-based multiple-information imaging (GB-MII) method, which can retrieve the attenuation, refraction and scattering information of samples by only three images. This algorithm aims at reducing the exposure time and the doses delivered to the sample. The multiple-information retrieval problem in GB-MII is solved by transforming a nonlinear equations set to a linear equations and adopting the nature of the trigonometric functions. The proposed algorithm is validated by experiments both on conventional X-ray source and synchrotron X-ray source, and compared with the traditional multiple-image-based retrieval algorithm. The experimental results show that our algorithm is comparable with the traditional retrieval algorithm and especially suitable for high Signal-to-Noise system.
Keywords :
x-Ray phase contrast imaging , Grating-based , Multiple-information retrieval
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2011
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2208131