Title of article :
Development of a thick gas electron multiplier for microdosimetry
Author/Authors :
Orchard، نويسنده , , G.M. and Chin، نويسنده , , K. and Prestwich، نويسنده , , W.V. and Waker، نويسنده , , A.J. and Byun، نويسنده , , S.H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
5
From page :
122
To page :
126
Abstract :
A new tissue-equivalent proportional counter based on a thick gas electron multiplier (THGEM) was developed and tested for microdosimetry. A systematic test was conducted at the McMaster Accelerator Laboratory to investigate the overall performance of the prototype detector. A mixed neutron–gamma-ray radiation field was generated using the 7Li(p,n) reaction. The detector was operated at low voltage initially to test the stability and then the relative multiplication gain was measured as a function of the operating high voltage. A drift potential of 100 V and a THGEM bias of 727 V generated a multiplication gain sufficient for the detection of both neutron and gamma-ray radiation. A consistent microdosimetric pattern was observed between the THGEM detector and standard TEPC for microdosimetry.
Keywords :
THGEM , THick Gas Electron Multiplier , microdosimetry , Tissue-equivalent proportional counter
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2011
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2208237
Link To Document :
بازگشت