Author/Authors :
La Rosa، نويسنده , , A. and Donetti، نويسنده , , M. and Borri، نويسنده , , M. and Rivero، نويسنده , , F. and Attili، نويسنده , , A. and Bourhaleb، نويسنده , , F. and Cirio، نويسنده , , R. and Garella، نويسنده , , M.A. and Giordanengo، نويسنده , , S. and Givehchi، نويسنده , , N. and Mazza، نويسنده , , G. and Marchetto، نويسنده , , F. and Pardo، نويسنده , , J. and Pecka، نويسنده , , A. and Peroni، نويسنده , , C.، نويسنده ,
Abstract :
An integrated 64-channel device for the read-out of parallel plate pixel and strip ionization detectors has been developed by the INFN and University of Torino. The detectors will be used for the monitoring and control of hadrontherapy beams. The ASIC has been designed in CMOS 0.8 μm technology and it is based on a current-to-frequency converter followed by a synchronous counter. In this paper, we present a detailed characterization of the device done with 113 chips.
Keywords :
radiotherapy , Hadrontherapy , VLSI electronics , CMOS technology , current-to-frequency converter , dosimetry , Beam position monitor