Title of article
Measurement of the non-linearity and interpixel capacitance of a near-IR detector
Author/Authors
Smadja، نويسنده , , G. and Cerna، نويسنده , , C. and Ealet، نويسنده , , A.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
7
From page
615
To page
621
Abstract
A temporal analysis of the noise of a H2RG detector is performed in view of the calibration of a dedicated setup. The implication on the noise performance of the subtraction of the reference output is described, and we extend the correlation method to groups of several pixels to derive the ratio x of the interpixel capacitance to the pixel capacitance of a detector. We find x = - 0.0259 ± 0.0017 ( stat ) ± 0.0020 ( syst . ) . All measurements are consistent to a sub-percent accuracy and provide a calibration of our setup.
Keywords
Poisson fluctuations , Noise correlations , shot noise , Readout noise , Calibration , Non-linearity
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2009
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2209136
Link To Document