Title of article :
Towards a magnetic field stabilization at ISOLTRAP for high-accuracy mass measurements on exotic nuclides
Author/Authors :
Marie-Jeanne، نويسنده , , M. and Alonso، نويسنده , , J. and Blaum، نويسنده , , K. and Djekic، نويسنده , , S. and Dworschak، نويسنده , , M. and Hager، نويسنده , , U. and Herlert، نويسنده , , A. and Nagy، نويسنده , , Sz. and Savreux، نويسنده , , R. and Schweikhard، نويسنده , , L. and Stahl، نويسنده , , S. and Yazidjian، نويسنده , , C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
10
From page :
464
To page :
473
Abstract :
The field stability of a mass spectrometer plays a crucial role in the accuracy of mass measurements. In the case of mass determination of short-lived nuclides with a Penning trap, major causes of fluctuations are temperature variations in the vicinity of the trap and pressure changes in the liquid helium cryostat of the superconducting magnet. Thus systems for the temperature and pressure stabilization of the Penning trap mass spectrometer ISOLTRAP at the ISOLDE facility at CERN have been installed. A reduction of the temperature and pressure fluctuations by at least an order of magnitude down to Δ T ≈ ± 5 mK and Δ p ≈ ± 5 Pa has been achieved, which corresponds to a relative magnetic field change of Δ B / B = 2.7 × 10 - 9 and 1.1 × 10 - 10 , respectively.
Keywords :
Penning Trap , Magnetic field , Precision mass spectrometry , Stabilization