Title of article :
Light collection optimization in scintillator-based gamma-ray spectrometers
Author/Authors :
Hull، نويسنده , , G. and Du، نويسنده , , S. and Niedermayr، نويسنده , , T. and Payne، نويسنده , , S. and Cherepy، نويسنده , , N. and Drobshoff، نويسنده , , A. and Fabris، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Scintillator-based gamma-ray detectors are being actively pursued for homeland security applications. A key property of such detectors is their energy resolution which enables faster detection and more precise identification of gamma-ray sources. In order to obtain the best energy resolution with a given scintillator material, it is crucial to collect the largest fraction possible of the light emitted after gamma-ray absorption. Different techniques to maximize the light collection efficiency were investigated and tested experimentally. In particular, the effect of the scintillator geometry has been simulated with Detect2000. Also, a number of wrapping materials have been tested for their reflectivity and their performance in terms of improving the energy resolution in a BGO-based gamma-ray detector. The best results were obtained with a tapered cylinder geometry and the GORE DRP tape.
Keywords :
Gamma radiation detectors , Scintillator detectors , Scintillators , Ray-trace modeling , BGO energy resolution , Diffuse reflectors
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Journal title :
Nuclear Instruments and Methods in Physics Research Section A