Title of article
Development of a pixel detector for ultra-cold neutrons
Author/Authors
Kawasaki، نويسنده , , S. and Ichikawa، نويسنده , , G. and Hino، نويسنده , , M. and Kamiya، نويسنده , , Y. and Kitaguchi، نويسنده , , M. and Komamiya، نويسنده , , S. and Sanuki، نويسنده , , T. and Sonoda، نويسنده , , S.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
6
From page
42
To page
47
Abstract
A pixel detector with high spatial resolution and temporal information for ultra-cold neutrons is developed based on a commercial CCD on which a neutron converter is attached. 10B and 6Li are tested for the neutron converter and 10B is found to be more suitable based on efficiency and spatial resolution. The pixel detector has an efficiency of 44.1 ± 1.1 % and a spatial resolution of 2.9 ± 0.1 μ m (1 sigma).
Keywords
UCN , Pixel detector
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2010
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2209687
Link To Document