Title of article
Analytical approach to calculate the efficiency of 4π NaI(Tl) gamma-ray detectors for extended sources
Author/Authors
Abbas، نويسنده , , Mahmoud I.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
5
From page
48
To page
52
Abstract
The 4π NaI(Tl) detectors are usually utilized for low-level radioactivity measurements of environmental samples in gamma-ray spectroscopy due to the near 4π solid angle and high full-energy peak and total efficiencies. An analytical theoretical approach is presented here and used to calibrate a 4π NaI(Tl) (2a×2b×2c) detector, with a central rectangle hole (2l×2m×2c) was left empty for sample access, for isotropic radiating (point, plane and volumetric) sources. The approach depends on the accurate calculation of two important factors; the path length d, the photon traverses within the active volume of a gamma detector, and the geometrical solid angle Ω, subtended by the source to the detector at the point of entrance. The comparisons with the experimental and Monte Carlo method works reported in the literature indicate that the present approach is useful in the efficiency calibration of such complicated gamma-ray spectrometer.
Keywords
Geometrical efficiency , Total efficiency , Extended sources , 4? NaI(Tl) gamma detectors
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Serial Year
2010
Journal title
Nuclear Instruments and Methods in Physics Research Section A
Record number
2209688