• Title of article

    Digital pulse processing for planar TlBr detectors

  • Author/Authors

    Nakhostin، نويسنده , , M. and Hitomi، نويسنده , , K. and Ishii، نويسنده , , K. and Kikuchi، نويسنده , , Y.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    3
  • From page
    242
  • To page
    244
  • Abstract
    We report on a digital pulse processing algorithm for correction of charge trapping in the planar TlBr detectors. The algorithm is performed on the signals digitized at the preamplifier stage. The algorithm is very simple and is implemented with little computational effort. By using a digitizer with a sampling rate of 250 MSample/s and 8 bit resolution, an energy resolution of 6.5% is achieved at 511 keV with a 0.7 mm thick detector.
  • Keywords
    digital signal processing , ?-ray spectrometers , Thallium bromide
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2010
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number