Title of article :
A thin emulsion spectrometer using a compact permanent magnet
Author/Authors :
Fukushima، نويسنده , , C. and Kimura، نويسنده , , M. and Ogawa، نويسنده , , S. and Shibuya، نويسنده , , H. and Takahashi، نويسنده , , G. and Kodama، نويسنده , , K. SRID HARA، نويسنده , , T. and Mikado، نويسنده , , S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
7
From page :
56
To page :
62
Abstract :
A 30 mm long spectrometer made of three emulsion sheets inside a compact permanent magnet has been tested with secondary beams produced by the proton synchrotron at KEK. A momentum resolution of σ ( p ) / p = 13.3 ± 0.3 % has been obtained in the momentum range from 0.5 to 2.0 GeV/c, using the sagitta measurements of the tracks reconstructed with an automated emulsion readout system.
Keywords :
Permanent Magnet , Nuclear emulsion , Emulsion Cloud Chamber , Magnetic spectrometer
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2008
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2209805
Link To Document :
بازگشت