Abstract :
A liquid scintillation detector aimed for neutron energy and fluence measurements in the energy region below 20 MeV has been calibrated using monoenergetic and white spectrum neutron fields. Careful measurements of the proton light output function and the response matrix have been performed allowing for the application of unfolding techniques using existing codes. The response matrix is used to characterize monoenergetic neutron fields produced by the T(d,n) reaction at low deuteron energies.
Keywords :
Liquid scintillator , Detector response , Unfolding , Neutron , Flux measurements