• Title of article

    Exploiting the MEDIPIX2 detector for the reconstruction of X-ray spectra

  • Author/Authors

    Talla، نويسنده , , Patrick Takoukam and Michel، نويسنده , , Thilo and Firsching، نويسنده , , Markus and Durst، نويسنده , , Jürgen and Anton، نويسنده , , Gisela، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    103
  • To page
    106
  • Abstract
    The single photon counting pixel detector MEDIPIX2 has been used in a variety of applications. Because of its spectral capability, the detector can be used as spectrometer. However, the measured spectrum is not the real incident spectrum as the MEDIPIX2 suffers from charge sharing because of its small pixel size.This implies that the energy information is smeared out heavily. Taking the measured spectrum and the responses to monoenergetic exposure we applied some deconvolution methods to determine the incoming spectrum.
  • Keywords
    x-ray imaging , photon counting detector , Pixel detector , Medipix2 , Spectrum reconstruction
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2009
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number