Title of article :
Validation of a computed radiography device to monitor the HIV-1 RNase H activity
Author/Authors :
Esposito، نويسنده , , F. and Fanti، نويسنده , , V. and Marzeddu، نويسنده , , R. and Randaccio، نويسنده , , P. and Tramontano، نويسنده , , E. and Zinzula، نويسنده , , L.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
3
From page :
226
To page :
228
Abstract :
A commercially available computed radiography (CR) system for dental radiography was used to produce images from radiolabeled polyacrilamide gel electrophoresis (PAGE) assays. Typically, similar investigations require specific and expensive autoradiography devices. The CR unit was characterized in terms of sensitivity and fading by means of a 90Sr source that well simulates the experimental conditions, and then used for quantitative analyses of the human immunodeficiency virus type 1 (HIV-1) reverse transcriptase (RT) polymerase-independent ribonuclease H (RNase H) activity monitored by PAGE analysis. The results showed that the present methodology allows quantifying effectively the RNase H catalyses and that the obtained data are in good agreement with previous reference works. Finally, in order to further validate the present method in terms of relationship between enzyme activity, the rate of products formation and signal intensity, a PAGE analyses of the HIV-1 RNase H inhibition by the known diketo acid derivative RDS1643 was carried out.
Keywords :
x-ray imaging , computed radiography , reverse transcriptase , HIV-1 , Antiviral agents
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2210022
Link To Document :
بازگشت