Title of article :
A single-photon CCD-based setup for in situ measurement of the X-ray spectrum of mammographic units
Author/Authors :
Labate، نويسنده , , L. and Levato، نويسنده , , T. and Galimberti، نويسنده , , M. and Giulietti، نويسنده , , A. M. Giulietti، نويسنده , , D. and Sanna، نويسنده , , M. and Traino، نويسنده , , C. and Lazzeri، نويسنده , , M. and Gizzi، نويسنده , , L.A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
5
From page :
278
To page :
282
Abstract :
A technique enabling in situ measurements of the spectrum of X-ray tubes employed in mammographic screenings is described. The technique involves the use of a commercially available CCD camera and a set of metal foils and is particularly useful to perform a fast evaluation of the spectral properties of a Mo anode mammographic system operating at standard flux levels. A description of the detector calibration procedure is first given, followed by a discussion of the study and choice of an appropriate set of X-ray attenuation foils. Finally, the use of the system for a spectroscopic characterization of a mammographic system is reported.
Keywords :
mammography , X-ray spectroscopy , quality control , CCD detectors