• Title of article

    On-line mirror alignment monitoring method for COMPASS RICH-1

  • Author/Authors

    Alexeev، نويسنده , , M. and Birsa، نويسنده , , R. and Bradamante، نويسنده , , F. and Bressan، نويسنده , , A. and Chiosso، نويسنده , , M. and Ciliberti، نويسنده , , P. and Colantoni، نويسنده , , M.L and Torre، نويسنده , , S. Dalla and Denisov، نويسنده , , O. and Diaz، نويسنده , , V. and Duic، نويسنده , , V. and Ferrero، نويسنده , , A. and Gayde، نويسنده , , J.Ch. and Giorgi، نويسنده , , M. and Gobbo، نويسنده , , B. and Kramer، نويسنده , , D. and Levorato، نويسنده , , S. and Maggiora، نويسنده , , A. and Martin، نويسنده , , A. and Menon، نويسنده , , G. and Panzieri، نويسنده , , D. and Pesaro، نويسنده , , G. and Polak، نويسنده , , J. and Rocco، نويسنده , , E. and Schiavon، نويسنده , , P. and Sozzi، نويسنده , , F. and Steiger، نويسنده , , L. and Sulc، نويسنده , , M. and Svec، نويسنده , , M. and Takekawa، نويسنده , , S. and Tessarotto، نويسنده , , F.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    3
  • From page
    194
  • To page
    196
  • Abstract
    COMPASS RICH-1 is a large size gaseous Cherenkov detector. Two reflecting spherical surfaces, covering a total area of about 21 m 2 hosted in the radiator vessel provide image focusing. The segmented surfaces are formed by 116 mirrors, with individual degrees of freedom for angular adjustment. An original method has been developed and applied for the on-line monitoring of the mirror alignment. Relative tilts between neighbouring mirror elements create discontinuities in the image of continuous lines reflected by the mirror wall. The measurement of the discontinuities can also provide information about the absolute tilt with a resolution of 0.1 mrad. An independent check and calibration of this method has been performed in laboratory using, as reference, a laser interferometer. sent the technical solutions adopted to implement the dedicated setup, the results of the method calibration by laser interferometer, and the preliminary results obtained measuring the alignment of the RICH-1 mirror wall.
  • Keywords
    Mirror alignment , COMPASS , rich
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2008
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number

    2210321