• Title of article

    Theoretical calculation on CR-39 response for radon measurements and optimum diffusion chambers dimensions

  • Author/Authors

    Askari، نويسنده , , H.R. and Ghandi، نويسنده , , Kh. and Rahimi، نويسنده , , M. and Negarestani، نويسنده , , A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    16
  • From page
    368
  • To page
    383
  • Abstract
    One method to measure radon gas concentration in the air with a long time of radiation is trace chemical etching technique. There is a direct proportion between the number of traces on solid-state nuclear track detectors (SSNTDs) and activity concentration of radon. In this paper, calibration constant for a cylindrical chamber with CR-39 detector has been measured analytically. Using this measurement, trace curves on the base of concentration for chambers with different heights and radii have been drawn. The results show that to measure radon gas concentration, the optimum chamber should have a height between 3.5 and 4 cm and a radius between 2.5 and 3.2 cm.
  • Keywords
    radon progeny , CR-39 detector , Activity concentration , Radon gas , Response
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Serial Year
    2008
  • Journal title
    Nuclear Instruments and Methods in Physics Research Section A
  • Record number