Title of article :
Electric and magnetic field optimization procedure for Penning trap mass spectrometers
Author/Authors :
Beck، نويسنده , , D. and Blaum، نويسنده , , K. and Bollen، نويسنده , , G. and Delahaye، نويسنده , , P. and George، نويسنده , , S. and Guénaut، نويسنده , , C. and Herfurth، نويسنده , , F. and Herlert، نويسنده , , A. and Lunney، نويسنده , , D. and Schweikhard، نويسنده , , L. and Yazidjian، نويسنده , , C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
7
From page :
635
To page :
641
Abstract :
Significant systematic errors in high-precision Penning trap mass spectrometry can result from electric- and magnetic-field imperfections. An experimental procedure to minimize these uncertainties is presented for the on-line Penning trap mass spectrometer ISOLTRAP, located at ISOLDE/CERN. The deviations from the ideal magnetic and electric fields are probed by measuring the cyclotron frequency and the reduced cyclotron frequency, respectively, of stored ions as a function of the time between the ejection of ions from the preparation trap and their capture in the precision trap, which influences the energy of their axial motion. The correction parameters are adjusted to minimize the frequency shifts.
Keywords :
Ion motion , High-precision mass measurements , optimization , Penning Trap