Title of article :
Analysis of time resolution in a dual head PET system using low pass filter interpolation and digital constant fraction discriminator techniques
Author/Authors :
Monzَ، نويسنده , , José M. and Lerche، نويسنده , , Christoph W. and Martيnez، نويسنده , , Jorge D. and Esteve، نويسنده , , Raْl and Toledo، نويسنده , , José Luis Gadea، نويسنده , , Rafael and Colom، نويسنده , , Ricardo J. and Herrero، نويسنده , , Vicente and Ferrando، نويسنده , , Néstor and Aliaga، نويسنده , , Ramَn J. and Mateo، نويسنده , , Fernando and Sلnchez، نويسنده , , Filomeno and Mora، نويسنده , , Francisco J. and Benlloch، نويسنده , , José M. and Sebastiل، نويسنده , , ءngel، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
4
From page :
347
To page :
350
Abstract :
PET systems need good time resolution to improve the true event rate, random event rejection, and pile-up rejection. In this study we propose a digital procedure for this task using a low pass filter interpolation plus a Digital Constant Fraction Discriminator (DCFD). We analyzed the best way to implement this algorithm on our dual head PET system and how varying the quality of the acquired signal and electronic noise analytically affects timing resolution. Our detector uses two continuous LSO crystals with a position sensitive PMT. Six signals per detector are acquired using an analog electronics front-end and these signals are processed using an in-house digital acquisition board. The test bench developed simulates the electronics and digital algorithms using Matlab ® . Results show that electronic noise and other undesired effects have a significant effect on the timing resolution of the system. Interpolated DCFD gives better results than non-interpolated DCFD. In high noise environments, differences are reduced. An optimum delay selection, based on the environment noise, improves time resolution.
Keywords :
Position sensitive detector , Positron emission tomography , Event time detection , Digital algorithms , system simulation
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2211253
Link To Document :
بازگشت