Title of article :
Detection efficiency calculation for photons, electrons and positrons in a well detector. Part I: Analytical model
Author/Authors :
Pommé، نويسنده , , S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
8
From page :
584
To page :
591
Abstract :
An analytical model is presented to calculate the total detection efficiency of a well-type radiation detector for photons, electrons and positrons emitted from a radioactive source at an arbitrary position inside the well. The model is well suited to treat a typical set-up with a point source or cylindrical source and vial inside a NaI well detector, with or without lead shield surrounding it. It allows for fast absolute or relative total efficiency calibrations for a wide variety of geometrical configurations and also provides accurate input for the calculation of coincidence summing effects. Depending on its accuracy, it may even be applied in 4π–γ counting, a primary standardisation method for activity. Besides an accurate account of photon interactions, precautions are taken to simulate the special case of 511 keV annihilation quanta and to include realistic approximations for the range of (conversion) electrons and β−- and β+-particles.
Keywords :
Modelling , photons , nuclear , Well detector , SIMULATION , efficiency
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Serial Year :
2009
Journal title :
Nuclear Instruments and Methods in Physics Research Section A
Record number :
2211292
Link To Document :
بازگشت