Ohmori، نويسنده , , H. and Mizutani، نويسنده , , M. and Kaneeda، نويسنده , , T. and Abe، نويسنده , , N. and Okada، نويسنده , , Y. and Moriyama، نويسنده , , S. and Hisamori، نويسنده , , N. and Nishimura، نويسنده , , N. and Tsunashima، نويسنده , , Y. and Tanaka، نويسنده , , J. and Kuramoto، نويسنده , , K. and Ezura، نويسنده , , A.، نويسنده ,
Abstract :
The purpose of this paper is to prove the reliability of artificial hip joints’ surface generated by the electrolytically treated grinding process; the evaluation was carried out in a hip simulator and animal tests. The evaluation indicated that the fabricated hyper-hemispherical shaped artificial hip joints presented a high quality smoother surface than 10 nm in Ra. The fabricated surface also showed its ability to prevent crystalline phase transition during the hip simulation test, suggesting that the stably modified layer was generated via grinding process. The results of animal test showed that the fabricated surface could have a highly-improved biocompatibility.